Microscopy - Study Notes
Introduction
- Microscopes are specialised optical instruments that produce magnified visual or photographic images of objects too small to be seen with the naked eye.
- Bright field and dark field microscopes were studied earlier; this chapter covers phase contrast and fluorescence microscopes and the electron microscope, which uses electrons instead of light.
2.1 Phase contrast microscope (PCM)
- Invented by Frits Zernike, a Dutch physicist; Nobel Prize in 1953. It allows observation of living cells.
- Principle: small phase changes in light, caused by differences in thickness and refractive index of parts of an object, are transformed into differences in brightness. The eye cannot detect phase changes but can detect brightness.
- Components: light source, condenser, objective and ocular as in a compound microscope, plus two special parts.
- Sub-stage annular diaphragm (phase condenser): placed below the condenser; creates a narrow, hollow cone of light and separates diffracted waves from direct waves.
- Phase plate (diffraction or phase retardation plate): at the back focal plane of the objective; one portion is coated with magnesium fluoride, the other absorbs light; it reduces the phase of incident light.
- Working: rays through regions of high refractive index are retarded; the phase difference between deviated and non-deviated rays is about one-fourth of the wavelength (\( \lambda/4 \)). Waves out of phase by one-half wavelength cancel, giving darkness.
- Applications: unstained living cells, highly transparent objects, intracellular components at high resolution, cell division, chromosomal and flagellar movements.
2.2 Fluorescence microscope
- Combines the magnifying properties of the light microscope with visualisation of fluorescence. A higher intensity (lower wavelength) light excites a fluorophore (fluorochrome), which emits light of a longer wavelength.
- Sir George G. Stokes, a British scientist, first described fluorescence in 1852; the shift towards a longer emission wavelength is the Stokes shift.
- Common fluorophores: DAPI (blue), FITC (green), Texas Red (red).
- Light source: xenon or mercury arc lamp; the mercury arc lamp is 10-100 times brighter than incandescent lamps and covers ultraviolet to infrared. Lasers or high-power LEDs serve complex techniques.
- Filters: excitation filter (illumination path), emission or barrier filter (imaging path), and dichroic beam splitter at 45 degrees between them.
- Working: ultraviolet light from a mercury arc lamp passes the exciter filter as blue light, is reflected by the dichroic mirror to the specimen, which emits green light; the barrier filter passes only green light to the detector.
- Fluorescence is called cold light because it does not come from a hot source.
- Diascopic fluorescence (transmitted light, dark field condenser) was shown by K. Reichert and O. Heimstadt in 1911. Episcopic fluorescence, the most common today, sends light through the objective, which acts as condenser and objective; quartz lenses are needed for deep ultraviolet.
- Applications: biomedical research and clinical pathology, multicolour staining, texture of coal, porosity in ceramics, identifying Mycobacterium tuberculosis.
2.3 Electron microscope
- Ultrastructure needs more than 10,000X magnification. Electrons have wavelengths of 0.001-0.01 nm against 400-700 nm for visible light.
- De Broglie (French) showed in 1924 that accelerated electrons behave like waves. Ernst Ruska and Max Knoll (German) invented the electron microscope in 1931; Ruska won the Nobel Prize in 1986. TEM was the first type.
- Types: TEM, SEM and STEM.
- Principle: electrons from the cathode of the electron gun are accelerated to the anode by high voltage; all lenses are electromagnetic.
- Specimen preparation: fixation with glutaraldehyde then osmium tetroxide, dehydration, embedding in epoxy resin, sectioning by ultra-microtome; TEM sections 50-100 nm on a copper grid stained with lead acetate, uranyl acetate or phosphotungstate; SEM samples on an aluminium stub.
- TEM parts: electron gun (tungsten filament, 50,000-100,000 volts), condenser lens, specimen stage, objective and projector lens, screen or CCD camera. Vacuum prevents scattering by gases; the sample sits in the middle of the column.
- SEM: first built by Knoll in 1935; 3-D surface images without sectioning; specimen fixed in liquid propane at -180 °C, dehydrated in alcohol at -70 °C and coated with platinum or gold; secondary and backscattered electrons detected; EDS detectors use silicon or germanium.
- STEM combines TEM and SEM principles and needs very thin samples.
- The foldscope, an origami paper microscope by Manu Prakash, magnifies 140X up to 2400X.